GBS Elektronik MCA166-USB User Manual
Page 45
Name
MMCA_SET_FAST
Description
The command sets the fast discriminator threshold.
1. Parameter
Threshold
integer value
0 ... 2499, default 400
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_SLOW
Description
The command sets the slow discriminator threshold.
1. Parameter
Threshold
integer value
0 ... 2499, default 400
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_PZC_MANUAL
Description
The command sets the PZC value and returns the PZC offset.
1. Parameter
PZC value
integer value
0 ... 2499
Result array
Byte offset 128
Number of measured pulses
integer value
Result array
Byte offset 130
averaged negative offset of
measured input pulses
integer value
Return Code
see MMCA_RESET command
Name
MMCA_SET_PZC_VALUE
Description
The command sets the PZC value.
1. Parameter
PZC value
integer value
0 ... 2499
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_PZC_TIME_OFFSET
Description
The command sets the time offset for the PZC measurement.
1. Parameter
Offset for low DTC
integer value
0 ... 31, default 15
2. Parameter
Offset for high DTC integer value
0 ... 31, default 15
Results
-
Return Code
see MMCA_RESET command
18 Please note that this function may cause a firmware hang up if the following requirements are not
meets:
- statistical distributed input pulses with a count rate between 500 and 30000 Cps
- the significant peak is located above the first quarter of the spectrum
19 The function requires a reference to a pzc result array, but the function has no effect on this array at
the moment.
45