GBS Elektronik MCA166-USB User Manual
Page 67

MCA Stabilisation Commands:
Command name
CMD_SET_STABILISATION
Description
The command sets the peak stabilisation.
fl:
0
turns stabilisation off
1
turns stabilisation on
rb+3 rb: Peak ROI begin (LLD<= begin < end) re: Peak ROI end (begin < end <= ULD, (end-begin) < 250) Command syntax Format integer integer integer long integer Parameter preamble command fl rb re end flag Byte String (HEX) A5 5A 4D 00 fl l fl h rb l rb h re l re h B9 9B Used setup value channels, lld, uld Affected setup value stab_state, stab_roi_begin, stab_roi_end MCA state STATE_READY Command name CMD_SET_STAB_PARAM Description The command sets the stabilisation parameters. st: time interval (default 10) sa: area (default 25000) Command syntax Format integer integer integer long integer Parameter Preamble command st sa end flag Byte String (HEX) A5 5A 67 00 st l st h sa l sa l sa l sa h B9 9B Used setup value - Affected setup value stab_time, stab_area MCA state STATE_READY 67