beautypg.com

A mca527 algorithm, formulas, Amca527 algorithm, formulas – GBS Elektronik MCA-527 User Manual

Page 67

background image

MCA527 Algorithm, Formulas

A

MCA527 Algorithm, Formulas

This appendix lists the algorithms used by the MCA for the following functions:

Area, Area uncertainty

Centroid

FWHM

Smooth

Strip

Energy Calibration

The Area and the Area Uncertainty Function:

The area algorithm calculates the number of counts above the background in a ROI. The background area is
determined by averaging 4 points on both sides of the peak (the ROI limit points and 3 outer points) and linear
fit:

where

Spectrum

i

= the absolute number of counts in channels i

ROI

begin

= the ROI’s start channel

ROI

end

= the ROI’s stop channel

i

= channel number

67

Background =

ROI

end

ROI

begin

1

8

i =ROI

begin

3

ROI

begin

Spectrum

i

i=ROI

end

ROI

end

3

Spectrum

i

Integral =

i= ROI

begin

ROI

end

Spectrum

i

Area Uncertainty=

Integral

i= ROI

begin

3

ROI

begin

Spectrum

i

i= ROI

end

ROI

end

3

Spectrum

i

ROI

end

ROI

begin

1

8

2

Area=IntegralBackground