Pulse Height Spectroscopical Measurements

3.2.9 Offset

Offset is normally set automatically. The internal offset DAC can be adjusted such that the
input range goes from 0V to +U

max

or from -U

max

to 0V. Typically it is set such for positive

signals the base line is at 10% of the ADC range and for negative signals the base line is
at 90% of the ADC range.

35

Figure 16: Dependence of resolution on flattop setting using a NaI detector. As a NaI

detector exhibits slower rise times, a bit longer flattops yield best results.

0

1

2

3

4

5

5,0

6,0

7,0

8,0

9,0

FWHM @ 663keV [%]

Flattop [µs]

F

W

H

M

@

66

2k

eV

[%

]

Figure 17: Spectrum with extremely wrong offset setting. The spectrum just cuts off

somewhere in the middle.