GBS Elektronik MCA-527 User Manual
Page 44

MCA527
3.2.17 MCA527 Setup Examples for Use With Different Detectors
This table shall just give an overview of possible settings useful with the MCA527. It is far
from being complete. For actual settings with your detector, see the detectors manual.
Table 12: Parameters for different detectors
Ty
p
e
M
o
d
e
l
M
a
n
u
fa
c
tu
re
r
H
V
[
V
]
12
V
-1
2
V
2
4V
-2
4
V
In
p
u
t
P
o
la
ri
ty
C
h
a
n
n
e
ls
S
h
a
p
in
g
T
im
e
[
µ
s
]
F
la
tt
o
p
T
im
e
[
µ
s
]
CdZnTe
SDP/Z/60
Ritec
+300
x
x
-
-
neg
1024
0.8
0.7
CdZnTe
SDP/Z/20
Ritec
+500
x
x
-
-
neg
1024
0.8
0.7
CdZnTe
CZT500
Ritec
+600
x
x
-
-
pos
1024
0.8
0.7
NaI
Scintiblock
Crismatec
+500
pos
512
0.2
1.6
NaI
25825
Scionix
+550
x
x
-
-
pos
512
0.2
1.6
NaI
40*40
Amcrys-h
-800
x
x
pos
512
0.2
1.6
CsI
Sc1010
Eurorad
-
x
x
-
-
neg
256
1
HPGe
GMX-20190-S
Ortec
-3000
x
x
x
x neg
4096
4
0.8
HPGe
GEM 15-190
Ortec
+3000
x
x
x
x pos
4096
4
HPGe
GR 2018
Canberra
-3000
neg
4096
4
HPGe
GL 0310
Canberra
-2000
x
x
x
-
neg
4096
2
Si planar 7865S
Ortec
-1000
neg
4096
HPGe
EGC 30-190R
Eurisys
-3000
pos
4096
4
HPGe
EGPC 25-185
Eurisys
+2000
neg
4096
4
Table 13: Further settings, optimized for different count rates,as example a coaxial HPGe
shaping
time
flat
top
trigger
filter
Base line
restorer
gain
PZC
low count rate (<4kcps),
optimization for best resolution and
dynamic range.
4µs
0.9µs -101
16
high coarse gain,
small fine gain
(0.5...1)
general purpose, medium count
rates 4-20kcps
2µs
0.8µs 10-201 8
high count rate and throughput,
input >20kcps,
1µs
0.7µs 1-21
4
lower coarse gain
to avoid overrange
adjustment
critical!
•
Pile up rejection should be always on, except there is a preamplifier signal with
overshoot and the pile up rejector rejects everything.
•
Jitter correction should be always off, its an experimental feature to improve
resolution.
•
LF rejection should be used if there is a problem with low frequency noise (e. g.
50Hz) microphonics (machinery nearby running) or similar. If using this, increase
shaping time by factor 3 at the same time.
•
Input polarity setting depends on detector; if no spectrum can be seen change this;
check with oscilloscope mode.
•
PZC should be always adjusted
44