Fluke Biomedical 37-705 User Manual
Page 38

ENERGY CONSTANCY REPORT
VeriDose V, Model number
37-705
Serial number 00000
Firmware revision 1.1 d, March 30, 1998
RADIATION ONCOLOGY
Date: 03/30/1998
Time: 17:15:33
Detector
Group:
GROUP
2
Cal Date (s):
03/30/1998
03/30/1998 03/30/1998 03/30/1998 03/30/1998
Diode S.N. (s):
1932
1933
1934
1935
1936
Machine:
LINAC LINAC LINAC LINAC LINAC
Modality:
electron electron electron electron electron
Energy:
006 MeV
006 MeV
006 MeV
006 MeV
006 MeV
Normalization/Reference Depth: 1.5 cm
Channel
1
Channel
2
Channel
3
Channel
4
Channel
5
DEPTH
(cm)
1.5 4.3 8.7 11.2 17.4
READING
(cGy)
96.7 112.6 93.9 76.8 38.5
NORMALIZED
(percent)
100.0 116.5 97.1 79.4 39.8
Signature:
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)