FLATNESS/SYMMETRY REPORT
VeriDose V, Model number
37-705
Serial number 00000
Firmware revision 1.1 d, March 30, 1998
RADIATION ONCOLOGY
Date: 03/30/1998
Time: 17:36:34
Detector
Group:
GROUP
2
Cal Date (s):
03/30/1998
03/30/1998 03/30/1998 03/30/1998 03/30/1998
Diode S. N. (s):
1932
1933
1934
1935
1936
Machine:
LINAC LINAC LINAC LINAC LINAC
Modality:
electron electron electron electron electron
Energy:
006 MeV
006 MeV
006 MeV
006 MeV
006 MeV
Field Size:
10.0 cm x 10.0 cm
Channel
1
Channel
2
Channel
3
Channel
4
Channel
5
MEASURED
DOSE (cGy)
109.6 109.9 110.2 109.8 110.5
POSITION
ASSIGNMENT
Center -X +X
-Y
+Y
RATIO 1.000
1.003
1.006
1.002
1.008
X-Axis Flatness:
0.274%
Y-Axis Flatness:
0.418%
X-Axis Symmetry:
0.242
Y-Axis
Symmetry: 0.618
Avg. Measured Dose:
110.0 cGy
Signature:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)