Fluke Biomedical 37-705 User Manual
Page 20

Nuclear Associates 37-705
Operators Manual
2-10
Password
Setup Password screen
MEASURE PRINTER
AAAA
AAAA
SELECT DAT/TIM
SETUP
QAPHANT
CAL
PASSWRD
NAME
DATA
MODE OPTION OLD NEW
This screen permits the operator to change the security password. The operator must enter the existing
password and a new password. The password is a 4 character, alpha-numeric entry. Press the OLD or
NEW soft-key to activate the column, and use the UP and DOWN keys to advance to the appropriate
character. Press the OLD or NEW soft-key to move the cursor to the next character position. Pressing the
ENTER key will save the password entry. The factory default password is AAAA.
Institution Name
Setup Name Screen
MEASURE PRINTER
AAAAAAAAAAAAAAAAAAAAAAA
SELECT DAT/TIM
SETUP
QAPHANT
CAL PASSWRD
NAME
DATA
MODE
OPTION
ENTER INSTITUTION NAME
This screen permits the operator to enter an institution name of up to 23, alpha-numeric characters.
Select the ENTER INSTITUTION NAME soft-key and use the UP and DOWN keys to advance to the
appropriate character. Press the ENTER INSTITUTION NAME soft-key to move the cursor to the next
character position. The institution name is printed on the reports.
Data Format
Setup Data screen
MEASURE PRINTER
US
SELECT DAT/TIM
EURO
SETUP
QAPHANT
CAL PASSWRD
NAME
DATA
MODE OPTION
STYLE
This screen permits the operator to select the data format as US or European. The US format provides
decimal points for the data while the EURO format provides commas in place of the decimal points. Press
the STYLE soft-key scroll to the appropriate format.