4 select mode – Fluke Biomedical 37-705 User Manual
Page 17

Operation
Measure Modes
2
2-7
Reports
Report Screen
MEASURE
DOSE TRTFULL
SELECT RATE
FLT/SYM
SETUP ALARMS
E-
CONST
CAL
REPORT
TRTLABL
MODE OPTION TYPE
The report screen is used to select the type of report to be printed. The selected report can be manually
or automatically printed following an exposure. Automatic printing is activated by turning the A-PRINT
function ON (ref. SETUP - PRINTER). To manually print a report based on the last exposure, select
report type and press the ENTER key.
There are four types of reports that can be printed directly from the VeriDose V. The default is the
previously selected report type.
• TRTFULL
Full treatment report
• TRTLABL
Treatment report for label printer only
• FLT/SYM
Flatness and symmetry report
• E-CONST
Energy constancy report
See the Appendices for sample reports.
2.4 Select Mode
Calibration Groups
Select Screen
MEASURE
group name
1 ON
SELECT
2
ON
SETUP
3
OFF
CAL
4 OFF
5
OFF
MODE GROUP CHANNEL
STATUS
MORE
In the SELECT - GROUP mode the operator can select a group of previously calibrated diodes. The UP
and DOWN keys may be used to scroll through the previously entered group names. The MORE soft-key
allows the operator to review the calibration data for each individual diode in the group. The operator
entries for MACHINE, MODALITY, ENERGY, COMMENT, STATUS, SER/NUM, CAL FAC, and DATE
(calibration date) are displayed via the MORE soft-key. When the channel STATUS is being displayed,
the highlighted channel can be toggled ON or OFF by pressing the STATUS soft-key. The highlighted
channel can be changed by pressing the CHANNEL soft-key.