Operation
Calibrate Modes
2
2-13
Calibrate Screen # 4
group
name
1
NO SER
NUM
AAAAAAA
2
Yes
MODALITY ELECTRON
3
ENERGY
000
4
MACHINE
AAAAAAA
5
SAVE CHANGES
TO group name?
GROUP CHANNEL
DIODE
YES
NO
After completing the entry of all the database information press the ENTER key. Press the YES key
(column 4) to save the database information.
Edit an Existing Group
Calibrate Screen # 5
MEASURE ADD
group
name
SELECT
EDIT
SETUP DELETE
CAL
EXPOSE
REPORT
MODE OPTION
GROUP
The EDIT option may be used to edit the database for the selected group. After selecting the EDIT option,
the operator may scroll through the existing groups by pressing the GROUP soft-key and pressing the UP
or DOWN key pads to the right of the display to scroll through the group names. After the group has been
selected press the ENTER key.
Calibrate Screen #6
group
name
NO
SER NUM
AAAAAAA
2
Yes
MODALITY
ELECTRON
3
ENERGY
000
4
MACHINE
AAAAAAA
5
COMMENT AAAAAAA
GROUP CHANNEL
DIODE
OPTION
After selecting the group to be edited, the items in the database are displayed. The operator may activate
each channel individually, assign a diode to that channel by selecting YES in the diode column, or de-
activate the channel by selecting NO. The database information regarding each diode in the set is dis-
played along the right side of the selected channel. The information is only displayed for active diodes
(indicated by YES in the diode column).
The SER/NUM entry allows the operator to record the serial number of the diode for the selected channel.
The MODALITY (modality) entries, selected by the scroll up and scroll down key, include ELECTRON,
PHOTON and CO60. The ENERGY entry allows the operator to enter the target energy for the selected
diode (the energy at which the diode is calibrated). The MACHINE entry may contain the name of the