Fluke Biomedical ESA612 User Manual
Page 88

ESA612
Users Manual
74
Differential leakage
Ranges ................................................................ 75 to 199
μ
A
200 to 1999
μ
A
2.00 to 20.00 mA
Accuracy .............................................................
±
(10 % of reading + (2 counts or 20
μ
A, whichever is greater))
Insulation resistance
Ranges ................................................................ 0.5 to 20.0 M
Ω
20.0 to 100.0 M
Ω
Accuracy
20 M
Ω
Range ..................................................
±
(2 % of reading + 0.2 M
Ω
)
100 M
Ω
Range ................................................
±
(7.5 % of reading + 0.2 M
Ω
)
Source test voltage ............................................. 500 V dc (+20 %, -0 %) 2.0
±
0.25 mA short-circuit current or 250 V dc selectable
Maximum load capacitance ................................ 1
μ
F
ECG Performance Waveforms
Accuracy
Frequency .......................................................
±
2 %
Amplitude ........................................................
±
5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle) ...................... 0.125 Hz and 2 Hz
Sine wave ....................................................... 10, 40, 50, 60, and 100 Hz
Triangle wave .................................................. 2 Hz
Pulse (63 ms pulse width) ............................... 30 BPM and 60 BPM