Using Memory
The non-volatile memory of the Analyzer will store up to
500 measurements or ECG information for each of 100
different Test Records. Each test record can be recalled to
the display of the Analyzer or exported to a PC. Data
Viewer software, available at
www.flukebiomedical.com/biomedical/usen/Support/softwa
re
, is required to upload memory data to a PC. Installation
and operator instructions are also located on the software
web page.
Note
It is recommended that memory data be
uploaded daily or as often as possible to reduce
upload time.
Storing Data into Memory
Each of the six measurement functions allow data storage
within the Analyzer. The ECG function allows storage of
the simulated waveforms only for reference of tests
performed. To store data, a Test Record ID should be
created first. To input a new Test Record ID:
1. Press
to open the screen shown in Figure 35.
2. Press
or to set the first character. Available
characters are 0-9, A-Z, and space.
3. Press
to move to the next character position.
4. Fill in up to 15 character positions to identify a test
record.
5. Press (enter save button) to store the Test Record ID.
6. Use
, , , and to enter the date of the Test
Record.
7. Press
to store the date.
The Test Record ID is displayed in the upper-left
corner of the display.
Note
If no record ID data is entered, a default ID of
0000000000000001 and a date of --/--/-- is
assigned to the record.