How to hold the product, Connecting to line power – Fluke Biomedical ESA612 User Manual
Page 24

ESA612
Users Manual
10
How to Hold the Product
When you move the Analyzer, use the handle in the
bottom case to hold it. See Figure 3.
Note
There is no handle on old versions of the
Analyzer.
fis122.eps
Figure 3. Product Handle
Connecting to Line Power
Warning
To avoid shock hazard and for proper
Analyzer operation, connect the factory
supplied three-conductor line power cord to
a properly grounded power outlet. Do not
use a two-conductor adapter or extension
cord; this will break the protective ground
connection.
Connect the Analyzer to a properly grounded three-prong
outlet. The Analyzer will not properly test a DUT when the
ground lead is open.
The Analyzer is intended for use with single-phase,
grounded power. It is not intended for dual, split-phase or
three-phase power configurations. But it can be used with
any power system that supplies the correct voltages for
single-phase and is grounded, or is an isolated power
system.