Performing a direct equipment leakage test – Fluke Biomedical ESA612 User Manual
Page 60

ESA612
Users Manual
46
To perform an alternative applied part leakage test:
1. Press
.
2. Press the soft key labeled More.
3. Select the desired applied part groupings using
and .
4. Press the soft key labeled Select.
5. Press the soft key labeled Alternative A.P..
6. Press
to apply the test voltage and read the
current in the display.
7. Press
or to advance to the next applied part
group(s) of a single function if applicable. Pressing
to read leakage current for each group.
Figure 26 shows the electrical connections between the
Analyzer and the DUT during an Alternative Applied Part
Leakage current test.
Note
If there are more than five applied parts to
connect to the Analyzer, see Using the 1-to-10
Adapter later in this manual.
Performing a Direct Equipment Leakage Test
Note
The Direct Equipment Leakage test is available
when the EN62353 standard is selected.
The Direct Equipment Leakage Current test measures the
leakage current between all applied parts and the exposed
conductive surface on the housing, to mains earth.
To perform a direct equipment test:
1. Press
.
The direct equipment test is the default test and should
already be selected.
2. Press
to apply the voltage and read the leakage
current in the display
Figure 27 shows the electrical connections between the
Analyzer and the DUT during a Direct Equipment Leakage
Current Test.