Connecting load cells to j-box, Changing setup and calibration numbers, Setup – Digi-Star EZ3400(V) User Manual
Page 38
Installation
D3714-GB Rev B EZ3400(V) User' s manual
34
Connecting load cells to J-Box
Figure4: J-Box connections
Changing setup and calibration numbers
[
SETUP.] & [CAL.]
1. Enter 871.
2. Press .
3. Weighing indicator briefly shows
SETUP
followed by a six digit number. Enter the
new number.
4. Press
.
Use 872 and follow the same procedure to
change the calibration
CAL number.
Connect load cell wires
to terminals.
Tighten nuts
Load cell cable
J-Box cable
Wire colours
Pin Colour
Description
1 Red
Power supply
+
2 Green
Signal -
3 White
Signal +
4 Black
Power supply
J-Box illustrated for
connection of 4 load cells
Connect to bottom panel of weighing
indicator. See figure 3.
1/3
2
SETUP
4
3