Menu descriptions
6100 0000 Instrument Type. Here you can change if the instrument will control or only measure mass flow or
pressure.
6110 0000 This will only let you measure the process. You can`t give setpoints.
6120 0000 This will let you control the process. You can edit the setpoint.
6200 0000 Sensor Type. Here you can select the used sensor type from the device.
This is mainly to obtain correctly direct reading in absolute values. Together with the capacity you
can define what 100% signal means. For example: sensortype ‘Gas volume’ gives readout units
like: ‘mln/min’, ‘ln/min’, ‘m3s/h’ etc.
After ‘UNIT’ the selected fluid number will be shown on the display.
6210 0000 Pressure sensor. This prepares module for readout in pressure units. In this mode the counter is
not available.
6211 0000 Select the pressure sensor readout unit with UP/DOWN key.
After unit selection choose: ‘a’ (absolute), ‘g’ (gauge), ‘d’ (differential) or ’. ‘ (no extension)
6220 0000 Liquid Volume Sensor. This prepares module for readout in liquid volume units.
6221 0000 Select the liquid volume readout unit with UP/DOWN key.
6230 0000 Liquid/Gas Mass Sensor. This prepares module for readout in mass flow units.
6231 0000 Select the liquid/mass flow sensor unit with UP/DOWN key.
6240 0000 Gas Volume Sensor. This prepares module for readout in gas volume (mass) units.
6241 0000 Select the gas volume readout unit with UP/DOWN key.
6250 0000 Other Sensor Type. This prepares module for readout in customized units. In this mode the
counter is not available, and you can not enter a capacity.
6251 0000 Here you can define your own sensor type. In order, you must enter the zero scale value (0%),
the full scale value (100%) and the unit type.
6300 0000 Sensor Capacity 0%. Here you can enter the capacity at 0%. This is the value which goes with
0% signal.
6310 0000 Edit the capacity at 0%.
6400 0000 Sensor Capacity. Here you can enter the sensor capacity. This is the value which goes with
100% signal at the selected readout unit.
6410 0000 Edit the sensor capacity.
6500 0000 Fluid Name. Here you can edit the fluid name (max. 10 characters).
6510 0000 Edit the fluid name.
6600 0000 Polynomial Constants. Here you can change the polynomial correction constants for this
instrument Normally these factors are obtained at calibration and represent a 3rd grade
polynomial function.
6610 0000 Here you can edit the polynomial constant A for selected fluidnr.
6611 0000 Edit constant A. (6520 to 6581 likewise for constants B...D or B…H, depending on the instrument
to be operate).
6700 0000 Fluidset Select. Here you can change the actual fluidset.
Your current settings will be saved and the settings of the selected fluidset will be the active
settings.
6710 0000 Select the active fluidset with UP/DOWN keys.
The fluidname will be displayed also (when filled-in).