Section 1. Features and Specifications
Electrical Characteristics

Electrical Characteristics

Switching Frequency

100 kHz (nominal)

Output Hold-up Time

25 ms at nominal line

Maximum Voltage Differential from
output to safety ground

±400 Vdc

Insulation Resistance

Input to chassis: >120
M

 Output to chassis:

>120 M



Isolation Voltage

(Output not to exceed ±400 Vdc from
chassis potential.)

Maximum Remote Sense Line Drop
Compensation.

(Line drop must be

deducted from the supply’s maximum
output voltage.)

Input to output: 1350 Vac

1.5 V/line

Environmental Specifications

Operating Ambient Temperature

0 to 30 °C with default local sensing. Above
30 °C, derate output linearly to 0 at 70 °C.

Storage Temperature Range

–55° to 85 °C

Humidity Range

0 to 80% RH non-condensing

Operating Altitude

Up to 6,500 feet (2000 m)

Storage Altitude

Up to 50,000 feet (15 000 m)

Installation Category

Intended for use in installation category
(overvoltage category) II (IEC 1010-1 standard)

Pollution Degree

Category 2 (IEC 1010-1 standard)













6