Operation
Accumulated Status, Status, and Fault Registers

48

Operating Manual for GPIB for XT/HPD Series Power Supply

Accumulated Status, Status, and Fault Registers

The GPIB option card uses three separate registers which are always active. They
are the accumulated status, status, and fault registers. You can use the status
commands shown in Table 3.8, “Status Commands” to activate the registers. The
bit register has eight conditions, each assigned a bit weight. When querying a
register, the controller returns a response which is the sum of the weights of all
relevant conditions.

Example:

The accumulated status register shows that PON, REM, CC and CV have all been
active since the last accumulated register query.

Table 3.10 shows the mnemonics and bit weights which correspond to each register
condition. You can control conditions in the fault and status register by using the
MASK and UNMASK commands.

Table 3.10 Accumulated Status, Status, and Fault Registers

Notes:

1.

Only CV, CC, OV, SD, ERR, PON and FOLD can be masked or unmasked.

2.

The error (ERR) bit is reset in the accumulated status, status, and serial poll registers with an
error query (ERR?).

3.

The accumulated status register is cleared with an accumulated status query (ASTS?).

4.

A fault is cleared with a fault query (FAULT?).

ASTS?

Query the Accumulated Status register.

ASTS 771

Controller response.

771 = 512 + 256 + 2 + 1 = PON + REM + CC + CV

Condition

Mnemonic

Bit Position Bit Weight

Constant voltage operation

CV

0

1

Constant current operation

CC

1

2

Not used

2

4

Overvoltage protection tripped

OV

3

8

Not used

4

16

Supply external shutdown active (J7-1)

SD

5

32

Foldback mode operation

FOLD

6

64

Remote programming error

ERR

7

128

Power ON (accumulated status, status
registers only)

PON

8

256

Remote mode (accumulated status, status
registers only)

REM

9

512