beautypg.com

AMETEK HPD Series User Manual

Page 34

background image

Operation
GPIB Operation

32

Operating Manual for GPIB for XT/HPD Series Power Supply

Serial Poll

In a serial poll, the controller polls each device.

Example:

The power supply responds with a 8-bit status byte defined as follows:

Table 3.2 Serial Poll Status Register

ibrsp

Return serial poll byte.

Bit
Position

Decimal
Weight

Description

Reset By

0 (LSB)

1

Fault - Set when any bit in the fault
register is set by a fault condition in the
supply. See also

“Accumulated

Status, Status, and Fault Registers”
on page 48

.

FAULT? query to reset

1

2

Not used

2

4

Not used

3

8

Not used

4

16

Ready - Set when power supply is
ready to accept commands.

Power supply, during
command processing
period

5

32

Error (ERR) - Set when ERR bit
asserted in status register. See also

“Accumulated Status, Status, and
Fault Registers” on page 48

.

ERR? query or a new
error-free command

6

64

Request Service (SRQ) - Set when
power supply requests service.

Serial Poll

7 (MSB)

128

Power On (PON) - Set when unit
initializes at power on.

CLR or Device Clear