MicroDust Pro Aerosol Monitoring System
5
Casella CEL
Figure 1.
Sampling probe
1.
Infra-red emitter
2.
Sample volume
3.
Infra-red detector
4.
Light stop
Under clean-air conditions, all light is prevented from reaching the received by
a light stop (4). When dust particles enter the sample volume (2), the beam of
infra-red light is scattered within a narrow angle (12° to 20°) to the receiver (4).
By using a narrow angle of scatter, the instrument is made less sensitive to
variations in the refractive index and colour of measured particulate.