Tsp and respirable adapters – Casella CEL Microdust Pro real time dust monitor User Manual
Page 38

MicroDust Pro Aerosol Monitoring System
38
Casella CEL
TSP and Respirable adapters
Figure 20.
Gravimetric and respirable adapter modules
1 Pump connection
8 Measuring chamber inlet
2 Adapter block
9 Locating spigot
3 Locating spigot
10 Cyclone
4 Inlet tube
11 Coarse dust container
5 Thumb screw
12 Filter cassette (25 mm / 37 mm)
6 37 mm-to-25 mm reducer
13 O-rings
7 MicroDust sampling probe
The basic design of the TSP and respirable adapters is the same, but with
different inlet fittings. The respirable adapter is a centrifugal separator that
uses the cyclone principle of dust separation. It draws only the respirable
fraction of particulate matter through the measurement chamber and onto the
filter (D50 cut = 4 µm at 2.2 litres/minute).