2 microphone specification (cel-252) – Casella CEL CEL-35X dBadge User Manual
Page 32
![background image](https://www.manualsdir.com/files/767539/content/doc032.png)
Casella CEL-35X dBadge Users Handbook – Page 32
Indication
Overload Indication:
140.4dB RMS (i.e. 0.1dB above the top of the
range),
Peak Overload Value: 143.4dB Peak Linear (i.e. 3.1dB above the top
of the range),
Stabilisation Time:
3 seconds after power up,
Operating Environment
Temperature:
0
o
C to +40
o
C,
Pressure:
65 - 108kPa,
Humidity ± 0.5dB:
30% - 90% (non-condensing),
Magnetic Fields:
Negligible,
Storage Environment
Temperature:
-10
o
C to +50
o
C,
Battery:
Internal NiMH, 28 hour battery life,
Charge Time:
Less than 90 minutes from discharged,
Dimensions mm (in): 72x47x52 (2.8x1.8x2.0),
Weight g (oz):
68 (2.4),
The
CEL-35X/IS
Intrinsically
Safe
dBadge
is
labelled
with the following:
ATEX:
I M1
II 1G
EEx ia IIC T2 (222
o
C)
EEx ia I
(T
a
= -20
o
C to +40
o
C)
7.2 Microphone Specification (CEL-252)
Classification:
Type 2
Nominal Open-Circuit Sensitivity (250Hz):
-28dB ± 3.0dB
(30mV/Pa)
Polarization voltage:
0V
Capacitance (polarized, 250Hz):
11-15pF
Operating temperature range:
0 to +40
o
C
Cartridge thermal noise:
25dB(A)
Dimensions:
IEC 61094-4 Type WS
2
FM/CSA:
Class 1
Division 1
Groups A, B, C and D.
Temperature classification T2