Cel-63x environmental & occupational noise meter, Broadband period result line – Casella CEL CEL-63X User Manual
Page 86
CEL-63x Environmental & Occupational Noise Meter
86
Casella CEL
If Ln‟s are also output then 5 functions are output immediately after the L
CPeak
,
103.2 above to give 11 functions in total e.g.
L
AF10
, 64.5, L
AF50
, 63.2, L
AF90
, 45.7, L
AF99.9
, 32.4
The date and time output is the date and time when that profile started.
NOTE
If the profile interval is set to 100ms then Normal Online mode will output a 1 second
profile to the RS232 output but save a 100ms profile to memory during the run.
Normal Period Result (output every 1 minute to 24 hours):-
If Period output is enabled then at the end of every period the meter will output
a broadband Period result line. If the 1/1,1/3 Octave L
eq
is enabled and the
meter has octave or third octave filters then an Octave Period result line is then
also output with 11 x Octave band L
eqs
or a third octave Period result line is
output with 33 x 1/3 octave L
eqs
. If two lines are output then the dates and times
will be identical and is the date and time of the start of the period that was
measured.
Broadband Period Result line:-
This is in format:-
functionN, valueN
The broadband functions are those selected in the current Setup- View (These
are the same functions which are displayed when recalling completed runs to
the CEL-63x LCD). Between 1 and 9 functions can be selected as well as 0 to 5
Ln functions giving up to 14 functions in total e.g.
AFmax
,102.4,L
AFmin
,34.3,L
C-Aeq
,-5.0,L
AF90
,65.4
NOTE
The Instantaneous SPL function (eg LAF) is never output. In Normal online mode.
NOTE
All Broadband Period function values are in dBs and all functions except LC-Aeq in
the range 0.0 dB to 199.9dB and LC-Aeq has a range -5.0 to 199.9dB.