2 cleaning and replacing the intake filter, 4 cleaning, Danger – BINDER FDL 115 User Manual
Page 56
![background image](https://www.manualsdir.com/files/767110/content/doc056.png)
FDL (E2.1) 02/2015
page 56/75
14.2 Cleaning and replacing the intake filter
The fresh air intake filter (fine particle filter for particles 1 µm up to 10 µm, class F6/EU6 acc. to EN
779:2002) at the top right-hand side must be cleaned or replaced from time to time, depending on the
degree of soiling. Pull out the slide and blow through the filter cartridge from the inside with compressed
air or replace it.
Intake filter
Slide
Figure 14: Fresh air intake filter
14.3 Cleaning and decontaminating the safety drying oven
Clean the unit after each use to avoid potential corrosion damage by ingredients of the charging material.
DANGER
Electrical hazard.
Danger of death.
∅
Do NOT spill water or cleaning agents over the inner and outer surfaces.
Before cleaning, turn off the unit at the main power switch and discon-
nect the power plug.
Completely dry the appliance before turning it on again.
14.4 Cleaning
Always keep clean the inner parts of the safety drying oven including the drip pans, drip trays
and exhaust-air ducts. Remove the remnants of coating material at suitable intervals.
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces
inner chamber
racks
door gaskets
Standard commercial cleaning detergents free from acid or halides.
Alcohol based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel
Standard commercial cleaning detergents free from acid or halides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge
parts
rear unit wall
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.