BINDER FDL 115 User Manual
Page 17

FDL (E2.1) 02/2015
page 17/75
 
•
Safety in case the technical ventilation fails 
The heating only turns on if the air circulation is already in operation. 
If the air circulation fails, the heating immediately turns off. In addition, there is a visual signal: red in-
dicator light “AIR” (3). As an additional indication there is an acoustical signal which can be reset on
the controller.
•
Moving parts participating in the work process 
It is impossible to touch the fan from outside the device or from the interior. 
•
Safety, measurement and control devices 
The safety, measuring, and control equipment is easily accessible via the housing top cover. 
•
Electrostatic charge 
The interior parts are grounded. 
•
Protection against touchable surfaces 
Tested according to EN ISO 13732-1:2008. 
•
Floors 
See operating manual chap. 3.4 for the oven’s installation. 
•
Ventilation 
Ventilation shall be realized by the operator according to GUV-R 500 chap. 2.29 “Verarbeiten von 
Beschichtungsstoffen” (Processing of paints) (for Germany).
•
Cleaning 
See operating manual chap. 14. 
•
Examinations 
The safety drying oven has been inspected by the “Deutsche Gesetzliche Unfallversicherung e.V. 
(DGUV), Fachausschuss Nahrungs- und Genussmittel, Prüf- und Zertifizierungsstelle im DGUV Test“
(German Social Accident Insurance (DGUV), Expert Committee: “Metal and Surface Treatment”, Test-
ing and Certification Body in DGUV) and bears the GS mark.