5 equipment and options – BINDER BF 53 User Manual
Page 44
BF (E1) 02/2015
page 44/51
Unit size
53
115
240
400
720
Electrical data
IP system of protection acc. to EN 60529
20
20
20
20
20
Nominal voltage (±10 %) 50/60 Hz
V
230 1N~ 230 1N~ 230 1N~ 230 1N~ 230 1N~
Nominal power
kW
0.40
0.40
0.68
0.85
1.25
Energy consumption at 37 °C / 98.6 °F Wh/h
11
20
33
53
80
Unit fuse 5 x 20 mm
230V / 10A / middle-time-lag (M)
A
10
10
10
10
10
external external external external external
Power plug
shock proof plug
Installation category acc. to IEC 61010-1
II
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
2
Electrical connection data BF-UL constructed acc. to cUL standard (for USA and Canada)
Unit size
53-UL
115-UL 240-UL 400-UL 720-UL
Electrical data
Nominal voltage (±10%) 60 Hz
V
115 1N~ 115 1N~ 115 1N~ 115 1N~ 115 1N~
Power plug
NEMA
5-15P
5-15P
5-15P
5-15P
5-20P
Nominal power
kW
0.40
0.40
0.68
0.85
1.25
Unit fuse 6,3 x 32 mm
250V / super-time-lag TT
A
12.5
12.5
12.5
12.5
16
external external external external external
Installation category acc. to IEC 61010-1
II
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
2
Legend: 1) without outer glass door 2) up to 98 % of the set value
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880 observing the recommended wall
clearances of 10 % of the height, width and depth of the inner chamber. Technical data refer to 100% fan
speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
12.5 Equipment and Options
To operate the incubator with forced convection BF, use only original BINDER accessories or
accessories / components from third-party suppliers authorized by BINDER. The user is
responsible for any risk arising from using unauthorized accessories.
Standard equipment
Microprocessor temperature controller with LED display and several time functions
Controller Timer functions: Delayed ON, delayed Off and temperature dependent delayed OFF
Temperature safety device class 3.1 acc. to DIN 12880 with visual temperature alarm
Adjustable ramp function
Rear exhaust duct, internal diameter 50 mm / 1.97 in with ventilation slide
Adjustable air change by rear exhaust duct (50 mm / 1.97 in ) with ventilation flap and front ventilation
slide
Four castors, 2 lockable (size 720 only)
RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer
output with RS 232/RS 422 interface converter