5. Press
button to return to normal display (actual value display) (automatically after approx. 30
seconds).
Adjust the temperature safety device following any changes of the set-point (chap. 7).
6.2 Display / entry of temperature and ventilation set-points (with selected
temperature ramp)
If previously a temperature ramp value has been selected (chap. 6.4.2):
Press button
in normal display / actual value display during ramp operation to have displayed the
actual ramp temperature set-point changing according to the selected gradient in addition to the entered
final set-points for temperature and fan speed.
The incubator is operating, the controller is in normal display (actual value
display). The
actual temperature value (example: 21.8 °C) is displayed:
1. Press
button
The display shows alternately “
SPr” and the actual temperature ramp set-point changing according
to the selected gradient (example: 42.7 °C):
This ramp set-point is only displayed, not adjustable.
2. Press
button
The display shows alternately “
SP” and the previous temperature set-point (example: 60 °C):
3. With the
buttons enter a set-point value between 0 and 100.
The desired temperature set-point can be selected in a temperature range from 5 °C /
9 °F above room temperature up to 100 °C / 212 °F.
Wait 2
seconds until the entered temperature value
is
taken over
(display flashing once).
4. Press
button to proceed to the fan speed entry.
The display shows alternately “
n” and the previous fan speed set-point (example: 100%):
5. Set the desired fan speed with the
buttons
The fan speed can be set to a value between 0% and 100%.
Wait 2 seconds until the entered value is taken over (display flashing once).