12.4 BF technical data
Unit size
53
115
240
400
720
Exterior dimensions
Width
mm / inch
634
24.96
834
32.83
1034
40.71
1234
48.58
1234
48.58
Height (incl. feet/castors)
mm / inch
617
24.29
702
27.64
822
32.36
1022
40.24
1528
60.16
Depth
mm / inch
575
22.64
645
25.39
745
29.33
765
30.12
865
34.06
plus door handle, and exhaust duct mm / inch
105
4.13
105
4.13
105
4.13
105
4.13
105
4.13
Wall clearance rear
mm / inch
100
3.94
100
3.94
100
3.94
100
3.94
100
3.94
Wall clearance side
mm / inch
160
6.30
160
6.30
160
6.30
160
6.30
160
6.30
Exhaust duct, outer diameter
mm / inch
52
2.05
52
2.05
52
2.05
52
2.05
52
2.05
Steam space volume
l / cu.ft. 70 / 2.47 142 / 5.02 283 / 10.0 457 /
16.15
808 /
28.55
Number of door(s)
1
1
2
2
2
Interior dimensions
Width
mm / inch
400
15.75
600
23.62
800
31.50
1000
39.37
1000
39.37
Height
mm / inch
400
15.75
480
18.90
600
23.62
800
31.50
1200
47.24
Depth
mm / inch
330
12.99
400
15.75
500
19.69
500
19.69
600
23.62
Interior volume
l / cu.ft. 53 / 1.9 115 / 4.1 240 / 8.6 400 / 14.3 720 / 25.7
Number of racks (regular / max.)
2/4
2/5
2/7
2/9
2/15
Load per rack
Kg / lbs
15 / 33
20 / 44
30 / 66
35 / 77
45 / 99
Permitted total load
Kg / lbs
40 / 88 50 / 110 70 / 155 90 / 199 120 / 265
Weight (empty)
Kg / lbs
43 / 95 64 / 141 104 / 230 145 / 320 180 / 397
Temperature data
Temperature range, 5 °C / 9 °F
above ambient up to
°C / °F 100 / 212 100 / 212 100 / 212 100 / 212 100 / 212
Temperature fluctuation
at 37 °C /
98.6 °F
≤ ± K
0.2
0.2
0.2
0.2
0.2
at 50 °C /
122 °F
≤ ± K
0.2
0.2
0.2
0.2
0.2
Temperature uniformity
(variation) 1)
at 37 °C /
98.6 °F
± K
0.4
0.3
0.3
0.4
0.4
at 50 °C /
122 °F
± K
0.7
0.6
0.8
0.9
0.6
Heating up time 2)
to 37 °C /
98.6 °F
minutes
12
22
12
18
21
to 50 °C /
122 °F
minutes
20
23
24
26
24
Recovery time after door
was opened for 30 sec 2)
at 37 °C /
98.6 °F
minutes
1
1
1
2
1
at 50 °C /
122 °F
minutes
1.5
2
2
4
4
Air change
at 50 °C /
122 °F
x/h
59
29
19
17
11