4 c 150 technical data – BINDER C 150 User Manual
Page 80

C 150 (E2) 12/2012
page 80/90
19.4 C 150 technical data
Exterior dimensions
Width
mm / inch
680 / 26.77
Height (incl. feet)
mm / inch
819/ 32.24
Depth
mm / inch
815 / 32.09
Depth plus door handle, I-triangle
mm / inch
54 / 2.13
Depth plus power supply connection and gas connec-
tion
mm / inch
60 / 2.36
Wall clearance rear
mm / inch
100 / 3.94
Wall clearance side
mm / inch
50 / 1.97
Number of doors
1
Number of inner glass doors
1
Interior dimensions
Width
mm / inch
500 / 19.69
Height
mm / inch
600 / 23.62
Depth
mm / inch
500 / 19.69
Interior volume
l
150 / 5.4
Number of shelves, series / max.
3/6
Size of shelf, width x depth
mm x mm
inch x inch
491 x 442
19.33 x 17.40
Weight (empty)
kg / lbs
95 / 209.5
Temperature data
Temperature range, 7 °C / 12.6 °F above ambient, up
to
°C / °F
50 / 122
Temperature fluctuation
± K
0.1
Temperature uniformity (variation) at 37 °C/ 98.6 °F
± K
0.4
Recovery time 1) after door open for 30 sec
at 37 °C/ 98.6°F
minutes
5
Humidity data
Humidity range
% r.H.
95 ± 2
CO
2
data
CO
2
range
vol.-% CO
2
0 to 20
Setting accuracy
vol.-% CO
2
0.1
Recovery time 1) after door open for 30 sec
at 5 vol.-% CO
2
minutes
7
CO
2
measurement
Drift-free CO
2
infrared absorption measurement
system
Connection hose nozzle DN6 for CO
2
Connection to the unit for hose with internal diameter
mm / inch
6 / 0.24
Electrical Data
IP system of protection acc. to EN 60529
IP
20
Nominal voltage (±10%) 50/60 Hz
V
230 1N~
Nominal power
kW
1.40
Energy consumption 2) at 37 °C/ 98.6°F
Wh/h
110
Power plug
Shock-proof plug
Installation category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Unit fuse
5x20mm / semi time-lag / 10 A