P. 15.5.3) – BINDER C 150 User Manual
Page 70
C 150 (E2) 12/2012
page 70/90
After a manual abortion, the C 150 reverts to its stand
ard operational status. The indication “ - - - “ in
the lower display shows that the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
If necessary, repeat the sterilization.
When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
Turn on the unit (chap. 5).
15.5.2 Aborting the hot-air sterilization by opening the outer door
For safety reasons, the sterilization cycle is automatically aborted if you open the outer door of the C 150.
Do not open the glass door and close the outer door immediately.
After an abortion caused by opening the outer door, the C 150 reverts to its standard operational sta-
tus. The indicat
ion “ - - - “ in the lower display shows that the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
If necessary, repeat the sterilization.
When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
Turn on the unit (chap. 5).
15.5.3 Aborting the hot-air sterilization using the main power switch
For safety reasons, the sterilization cycle is aborted if the C 150 is turned off or if there is a power failure.
Normally you should not terminate sterilization in this way.
Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C / 98.6°F.
After turning on again the C 150, it reverts to its s
tandard operational status. The indication “ - - - “ in
the lower display shows that the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
If necessary, repeat the sterilization.
When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
Turn on the unit (chap. 5).