Caution – BINDER C 150 User Manual
Page 69

C 150 (E2) 12/2012
page 69/90
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting steri-
lization.
Aborting sterilization after less than 6 hours
When aborting the sterilization prematurely, it may be that the cells/pathogens inside the unit have not all
been killed. You should repeat the sterilization.
CAUTION
Interruption of temperature reaction time.
Ineffective sterilization.
Repeat the sterilization.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
Do NOT touch the glass door and inner surfaces for approx. 7 hours after aborting steri-
lization.
When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
Turn on the unit (chap. 5).
Three events which can abort the automatic sterilization cycle prematurely:
A manual abortion (chap. 15.5.1)
Opening the outer door (chap. 15.5.2)
Turning off the power to the C 150 or a brief power failure (chap. 15.5.3)
15.5.1 Aborting the hot-air sterilization manually
You can abort the sterilization cycle manually.
Press the controller button “180 °C” for three seconds during the sterilization cycle.
The lower display will show the operating function “StE”, and the value “ON” will appear in the upper
display.
Press “▲”.
The upper display toggles from “ON” to “OFF“.
Press “MODE” to end the sterilization cycle.
Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C / 98.6°F.