Starrett SR200 Surface Roughness Tester User Manual
Page 22

5.2
SR200
Specifications subject to change
Cut off
A few trial measurements made on different surfaces will soon demonstrate that on some, the
results obtained are very dependent upon the cut-off selected. This shows that it is important
to choose the cut-off to suit the surface. In general, fine surfaces require short cut-offs and
rough surfaces a longer one. The table on the following page gives some guidance on suitable
cut-offs.
Evaluation Length
A long evaluation length is mainly for use on sheet metal and similar materials, where a longer
surface is required to be representatives of the material being examined.
Table of Cut-off values
If not otherwise indicated on a drawing, the following should be used to determine the cut-off
λc (ISO4288)-1996. Recommended Cut-off ISO 4288-1996
Periodic Profiles
Non-periodic Profiles
Cut-offs
Sampling	Length/
Evaluation	Length
Spacing	Distances		
Sm	(mm)
Rz(µm)
Ra (µm)
λc (mm)
λc/L (mm)
>0.013	to	0.04	
>0.04	to	0.13	
>0.13	to	0.4	
>0.4	to	1.3	
>1.3	to	4
(0.025)	to	0.1	
>0.1	to	0.5	
>0.5	to	10	
>10	to	50	
>50	to	200
(0.006)	to	0.02	
>0.02	to	0.1	
>0.1	to	2	
>2	to	10	
>10	to	80
0.08	
0.25	
0.8	
2.5	
8
0.08/0.4	
0.25/1.25	
0.8/4	
2.5/12.5	
8/40
Operating Error Indications
During	a	measurement	a	message	may	be	displayed	which	indicates	that	an	error	condition	has	
occurred.		The	messages	and	the	probable	causes	for	their	display	are	as	follows:	
Display
Reason for message
Motor Error
Motor fault
Pick-up Error
Response at pick-up level error or faulty connection
Over range
Selected range too small
E
Can	occur	in	front	of	a	measured	value	due	to	excessive	slew	rate	for	
a	single	data	point	(large	spike).		May	also	occur	if	the	pick-up	is	lifted	
off	the	surface	
Data transmit error
Printer	connected	at	start	of	transmission	but	later	detected	as	not	
connected	(printer	off	line)	
Battery low
When	battery	voltage	<6.4V:	key	push	operations	are	disabled	and	
display	is	off.	
When	the	instrument	is	in	the	sleep	mode	and	the	MEASURE		key	
is	pressed,	then	“Battery	low”	is	displayed	for	2	seconds,	followed	by	
the	normal	display	or	operation.	
Stored surface profile is erased
Renew or recharge battery