The default settings are – Starrett SR200 Surface Roughness Tester User Manual
Page 14

3.2
SR200
Specifications subject to change
Making a measurement
Note 1: If the pickup has been changed or the instrument is being used for the first time, the instrument
should be calibrated (see Chapter 7).
Note 2: Successful use of the SR200 will only be possible if it is operated on a surface free from external
vibration - see also operating notes in Chapter 5.
Switching the SR200 ON
Pressing	the	SCROLL		key	brings	the	display	on	and	the	previously	selected	set-up	is	
displayed	(provided	power	has	been	continuously	present).		The	display	is	automatically	turned	
off	if	the	instrument	is	not	used	for	30	seconds.	
If	the	battery	is	dead	or	has	been	removed,	the	previously	selected	set-up	is	lost.		When	a	
battery	has	been	replaced	and	the	SCROLL		key	is	pressed,	the	start	up	message	will	display	
for	2	seconds	and	then	the	default	settings	are	restored.
The default settings are
Parameter:
Ra
Cut-off:
0.8mm
Evaluation length:
4.0mm
Range:
100mm
Data dump evaluation length:
4.0mm
Data dump range:
100mm
No	parameter/graph	selected	for	printout
Language	
English
Filter
Gaussian
If	the	user	wishes	to	change	any	of	these	settings,	this	can	be	carried	out	using	the	SCROLL		
and	SELECT		keys	(see	Chapter	4	for	more	detail)	
Press	the	MEASURE		key.		When	the	measurement	is	completed	the	pickup	returns	and	the	
results	are	displayed	on	the	screen	(see	example	below):	
GaussLc
0.80mm
Eval	Length 25.0mm
Ra
= 0.00µm
Rz
= 0.00µm
Rt
= 0.00µm
Rp
= 0.00µm
RSmm
= 0.00µm
more...>