4848B Reactor Controller
w w w . p a r r i n s t . c o m
23
MTM Module
Press return key and release
Operation
Mode
Select
type/value Comment
r-S
Run
Run/Stop
SP
1
Decimal point position
LoC
OFF
Lock mode (lock all keys or
only up/down arrow able to
use)
Press and hold down “SET” for 5-sec
Operation
Mode
Select
type/value Comment
InPt
v10
Input type (v5 = 0-5V, v10 =
0-10V, nA0 = 0-20mA, nA4 =
4-20mA, nV =0-50mV)
tP-H
To be deter-
mined**
Upper-limit range
tP-L
0
Lower-limit temperature
range
CTRL
ON/OFF
Control mode (ON/OFF,
MANUAL, PID and PID
PROG)
S-HC
Heat
Heat/Cool control
ALA1
0
Alarm mode
SALA
OFF
System alarm
CoSH
ON
Communication write func-
tion that able to use set
point from software
C-Sl
RTU
Format type
C-no
*
Communication address
bPS
9600
Communication baud rate
Len
8
Data length setting
PrtY
Even
Parity bit setting
StoP
1
Stop bit setting
* See note on address at end of Factory Defaults
** For 90VDC motor, 115VAC, tP-H = 142.9
For 180VDC motor, 230VAC, tP-H = 120.4
For 90VDC motor, 230VAC, tP-H = 153.2
Press “SET” and release
Operation
Mode
Select
type/value
Comment
HtS
0
Hysteresis
TPoF
0
PV Offset
CrHI
0
Analog output high
CrLo
0
Analog output low
* Alternate C-no Defaults for 4848B
The C-no Address in a 4848B takes special con-
sideration. Each module must have a unique c-no
address to properly communication with a PC. For
the 4848B assign the c-no address according to the
following diagram.
C-no=1
C-no=2
C-no=3
C-no=4
C-no=5
C-no=6
C-no=7