Martin Christ Epsilon 2-6D LSCplus User Manual
Page 106
![background image](/manuals/739289/106/background.png)
Freeze-dryer EPSILON 2-6 LSCplus
10 Technical data
106
Version 10/2012, Rev. 1.6 of 16/12/2014 • sb
Translation of the original operating manual
Connection requirements
Electrical connection:
3 x 400 V / 50 Hz
(other supply data on request)
Protection class:
I
IP-Code according to DIN 60529:
21
Nominal power:
3.45 kVA
Nominal current:
8.0 A
Mains fuse:
12 A
Equipment connections
Vacuum connection:
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Media drain valve:
Hose nozzle DN10 (outside diameter 12 mm)
Aeration valve:
Hose nozzle DN6 (outside diameter 8.5 mm max)
Mains input:
Cekon CEE (16 A 3ph)
Data interface (LAN):
RJ 45
Special equipment: water cooling system
Part number:
113120
Water feed flow temperature:
+15°C to +20°C
Water return flow temperature:
25°C max.
Water temperature difference:
10K max. with +15°C feed flow temperature
5K max. with +20°C feed flow temperature
Cooling water pressure:
10 bar max.
Differential pressure between feed flow and return
flow:
1.5 bar max.
Volumetric flow rate:
0.38 m
3
/h max.
Heat dissipated by cooling water:
3.065 kW
Cooling water feed flow connection:
R3/4" with hose nozzle DN 13
Cooling water return flow connection:
R3/4" with hose nozzle DN 13
Option: cooling water flow controller (factory setting):
12 bar (refrigeration unit, high pressure side)
Pos: 233 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1