beautypg.com

Laerdal VitalSim User Manual

Page 26

background image

26

Blood Pressure:
Pressure range

0 - 300 mmHg

Accuracy

+/- 4 mmHg

Calbration:

Pressure sensor must be calibrated to sphygmomanometer. See Setup menu.

Pulse:
Available pulses:

Carotid, Brachial and Radial
Only 2 pulses can be palpated simultaneously. (Power considerations).

Palpated BP:

Radial pulse turns off at systolic BP.
Brachial pulse turns off at 20 mmHg, to prevent noise generation in
auscultation area.

Adult

Child

Infant

Manikin

Manikin

Manikin

Default Blood Pressures (mmHg)

120/80 100/70

94/66

Pulses are adjusted to BP as follows:

Systolic BP

Carotid Pulse

Radial/Brachial Pulse

>= 88

Normal

Normal

< 88

Normal

Weak

< 80

Normal

Absent

< 70

Weak

Absent

< 60

Absent

Absent

Accessories

200-11350

Base Unit Battery Case

200-11450

Remote Control Battery Cover

200-11250

Remote Control to Base Unit Cable

200-10250

USB Cable

200-10550

DC Power Supply

200-12050

Direction for Use

200-11150

Operating Software on CD-Rom

200-10150

Carry Case

New VitalSim DfU En 6/14/05 4:18 PM Side 26