Maintenance – Laerdal SimJunior Standard User Manual
Page 21

19
Maintenance
Maintenance After Each Simulation Session
The following preventive measures are required to ensure
longevity of the SimJunior manikin.
− Return manikin and PC to their original condition.
– Power off the SimPad, Link Box and PCs.
– Charge batteries if necessary.
– Flush the IV-arm system and drain the IO leg (leave the plug in the
bottom of the IO leg open).
– After each session where the IV-arm has been used, flush the IV
arm with 60% Isopropanol alcohol or 70% ethanol.
– Wipe the skin with a moist cloth.
– Remove wet clothes or linens.
– Remove tape residue with a moist cloth or rubbing alcohol
– Based on the use of the manikin, replace modules that are
consumed or damaged.
– Replace cables or connectors showing visible damage.
– Servicing should be performed by qualified personnel at regular
intervals.
Always perform a service:
− If liquids have been spilled in or on the manikin.
− After use in dusty environments.
For information on and spare parts, see section: Spare Parts
and Accessories.
Before Storage or Shipping
– Flush the IV arm with 60% Isopropanol alcohol or 70% ethanol.
Open the drain valve on the bottom of the manikin’s right foot.