beautypg.com

Labconco Guardian 500 Airflow Monitor User Manual

Page 5

background image

5



External Connections
-- the alarm unit will have the following connection points :-



External Alarm --- volt free relay input ( close contact to activate the input )

Alarm Disable --- volt free relay input ( close contact to activate the input )

Sash High Input: a. Connection point for Sash High micro switch.
( Switch contact to close and remain closed in
Sash High condition )

b. Connection point for Sash High proximity switch.
( Switch contact to close and remain closed in
Sash High condition )


Note:- Use input a. OR input b. for the Sash High alarm



Output 1
--- volt free relay output for Low Air Alarm



Power supply --- low voltage DC power supply