Labconco FreeZone Clear Stoppering Chambers User Manual
Page 25

21
Product Service: Domestic 1-800-522-7658, International 816-333-8811
C
C
H
H
A
A
P
P
T
T
E
E
R
R
5
5
M
M
A
A
I
I
N
N
T
T
A
A
I
I
N
N
I
I
N
N
G
G
Y
Y
O
O
U
U
R
R
C
C
L
L
E
E
A
A
R
R
S
S
T
T
O
O
P
P
P
P
E
E
R
R
I
I
N
N
G
G
C
C
H
H
A
A
M
M
B
B
E
E
R
R
Under normal operation, the Clear Stoppering Chamber requires
little maintenance. The following maintenance schedule is
recommended:
As needed:
1. The user has the responsibility for carrying out appropriate
decontamination if hazardous material is spilled on or inside
the equipment. This may be done by wiping the contaminated
surfaces with a soft cloth dampened with alcohol. Alcohol
may craze the acrylic parts. Before using any cleaning or
decontamination method except those recommended by
Labconco, users should check with Labconco that the proposed
method will not damage the equipment.
2. Clean up all spills; remove liquids from the chamber.
3. Check oil level of the vacuum pump. It should be between
MIN and MAX. If the oil level is less than an inch (25.4 mm)
above MIN, add oil to proper level.
4. If oil shows cloudiness, particles or discoloration, drain the
pump and replace with fresh oil.
5. Utilization of acids requires immediate cleaning and
neutralization after a run or physical damage will result.