Labconco CentriVap micro IR Vacuum Concentrators 77010 Series User Manual
Page 25
Chapter 7: Troubleshooting
Product Service 1-800-522-7658
21
PROBLEM
CAUSE
CORRECTIVE ACTION
Evaporation rate is
reduced (continued)
Obstruction in hose
Remove obstruction or replace hose.
Lack of adequate vacuum See below.
Temp set point too low
Increase temperature set point.
No vacuum/poor vacuum
Liquid in glass trap
Empty trap.
Leaks in lines or
connectors or gasket
Locate and repair.
Foreign material on lid
gasket
Clean gasket and lid.