Refer to the following if your CentriVap micro IR fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.
CAUTION: Disconnect power before corrective action is taken.
PROBLEM
CAUSE
CORRECTIVE ACTION
Unit will not operate
Unit not connected to
electrical power
Connect unit to proper electrical receptacle.
Circuit breaker blown
Correct electrical problem and reset circuit
breaker.
Lid open
Close lid.
Excessive vibration
Sample vials not located
symmetrically in rotor
Reposition sample vials.
Sample odor in lab
Vent hose exhausting into
lab area
Redirect hose to fume hood.
Evaporation rate is
reduced
Heater inoperable
Contact Labconco.
Vacuum pump failure
Check pump.
Liquid in vacuum line
Clear vacuum line