Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 44204 Series User Manual
Page 73

67
Wash 1, Steam, Rinses 1 & 2, Dry Time +/- 1 minute
Wash 2, Final Rinse Time after reaching SP +/- 1 minute
Conductivity varies between 100 - 1200
Confirm operation of low detergent and acid rinse monitor
Detergent Dispense 18 - 22 ml
Rinse Aid Dispense 5 - 7 ml
5
Observe Wash Action – Water hits top of tank from spindles
INSP/TESTS No. CONDITIONS
AND
REQUIREMENTS
INST/LCC# RESULTS PASS
6
Run rinse only to wet surfaces.
Visual
After wash pump starts, place a wet towel across the overflow contacts.
Wash pump should stop and the washer should drain. Display should
show ALARM OVERFLOW.
Hi-Pot
1
1000V for 1 minute.
Hi Pot Tester ____________
Final
1
Check for Leaks
Visual
2
Unit must be Clean and Dry
Visual
3
All parts in place in good condition
Visual
4
Verify that thermostat on back of tank is installed properly.
Visual
5
Make sure drain hose is routed over top of wash pump.
Visual
6
Perform FACTORY RESET
Visual