Glassware washer/test report – Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 44204 Series User Manual
Page 67

61
Glassware Washer/Test Report
MODEL NUMBER
VOLTS/HZ
M.O. NUMBER
SERIAL NUMBER
INSP/TECH/
DATE
QTY/OF
INSP/TESTS No. CONDITIONS
AND
REQUIREMENTS
INST/LCC# RESULTS PASS
Set-Up
1
Connect tap and pure water.
2
Connect
drain.
3
Connect proper electrical power. (Use 115V, 60 Hz or 230V, 50 Hz)
4
Connect current probe to power line.
5
6
7
Close detergent cup door
Connect computer via RS232
Set up washer control
A. Open door
B. Press and hold ▼ button- Simultaneously close and latch door.
C. Press ▼ button to select function. Press ◄ or ► to actuate
function.
1) Set proper voltage
2) Set Temp Units - °C
3) Set DI Pump to Enable- YES
4) Set Light - ON- (window models only)
5) Set Liquid Deter Opt- NO
6) Set Factory Reset – YES
D. Press RUN
Hi-Pot
1
1000V for 1 minute.
Hi Pot Tester
Temperature
1
Position thermocouple near temperature probe.
Test
2
Press and hold ◄ button - simultaneously close and latch door.
3
Display must agree with temperature meter ±2°C.
Temperature Meter
Self Test
1 Open
door
2
Press & hold ▲ button – Simultaneously close and latch door.
3 Press
▼ button to select function. Press ◄ or ► to actuate function.
TAP
WATER
Visual
PURE
WATER
Visual
Continue in this mode until Float Switch shuts off water flow.
Visual
Open door to check water level. Bottom of heater must be
partially wetted. Top of heater must be partially unwetted.
Close door and resume DIAGNOSTICS.
WASH
PUMP
Visual