3.5 QUICK / FULL BUMP TEST
This stage of the bump test is dependent on whether QUICK
or FULL bump is configured.
3.5.1 Quick Bump Test
As the alarm threshold for each range is exceeded, the
audible / visual / vibration alarms will activate and a
symbol will appear, otherwise a
symbol will be
displayed.
3.5.2 Full Bump Test
After a short period of time, the gas readings are checked
against configurable limits. The audible / visual / vibration
alarms will activate and a
symbol will appear if the
readings are within these limits, otherwise a
symbol
will be displayed.
3.6 ALARM CONFIRMATION
The user is then prompted to confirm if the audible and
visual alarms were activated, as shown in Fig. 3-4
Fig. 3-4 Confirm Alarms (4-gas model)