1-5

INTRODUCTION

Motion sensor to detect movement and activate alarms
in a ‘man down’ situation.

Alphanumeric display with screen backlighting.
Backlighting is coloured green during normal
operation and red during alarm condition.

Internal electric pump (optional) with a nominal flow rate
of 0.5 to 0.7 litres per minute.

Both manual and fully automatic data logging.

Powered by an internal Li-Ion (Lithium Ion) rechargeable
battery, this will provide an operating time of up to 14
hours (non-pumped) or 8 hours (pumped). Maximum
recharge time is 4 hours.

Robust crocodile clip to allow fitting to belt, pocket, etc.

Communications interface to allow downloading of
stored data.

Fully certified to international standards.

Comprehensive range of accessories available.

‘TYPICAL OPERATING PARAMETERS’ are detailed in
Appendix ‘A’ of this Handbook.

1.3 DATA LOGGING

Data logging is a standard feature of all PS200 series
instruments and allows gas measurements, event logs, bump
tests and calibration details to be automatically stored and
later downloaded to a Personal Computer (PC) via a USB
connection.

The instrument can store in excess of 24 hours of readings
at a recording interval of 1 minute, 180 On / Off event logs
and alarms, 180 bump test logs and 8 calibrations.