6 alarms reset or acknowledge, 1 confidence signal, Acknowledge – Heath Consultants Personal Surveyor (PS200 Series) User Manual
Page 40: Alarms reset or acknowledge, Confidence beep, Confidence signal, Reset, alarms, Signal, confidence
2-16
USER HANDBOOK
2.6 ALARMS RESET OR ACKNOWLEDGE
When the instrument detects an alarm set point has been
reached, the audible, visual and vibrating alarm will be
activated to alert the user.
The alarms are individually programmable to be either
Latching, (i.e. alarm will stay on until the user resets by a
press and hold of the Right Hand (RH) button
when
the gas reading has returned within the preset alarm limits),
or Non-Latching (i.e. the audible and visual alarm will reset
automatically when the reading returns within the preset alarm
limits). Refer to Alarms Table, Chapter 4, for individual alarms.
Note:
Default alarms are set in accordance with current
international standards.
The audible alarm on each pre-set alarm can be muted for a
period of 60 seconds by a press and hold of the Right Hand
(RH) button
. After this period, should the gas value
remain outwith the pre-set alarm limit, the non-latching audible
alarm will become active again. If latching, the audible alarm
will become active again regardless of gas value.
2.6.1 Confidence Signal
During normal operation, the instrument sounds a confidence
beep and illuminates the green LED’s briefly every 15
seconds. This function makes the user aware that the
instrument is operating correctly:
Note: The confidence beep and / or LED’s can be disabled.
Refer to the optional ‘CONFIGURATION HANDBOOK’ CD-
ROM (Part No. 64193) for further information.