• Follow LCD calibration indication. Adjust the calibration knob

until the display shows “

12.88 mS” i.e. the conductivity reading

@ 25°C.

• All subsequent measurements will be compensated to 25°C (77°F).

If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), adjust the knob to read

11.67 mS” (see the conductivity vs. temperature chart on page 13).

All subsequent measurements will be compensated to 20°C.

• The calibration is now complete and the instrument is ready for

use.

The instrument should be recalibrated at least once a month, or
when the probe is changed.

Note: For more accurate results, it is advisable to use a calibration

solution close to the range to be measured. See the “Accessories”
section for a wide selection of conductivity solutions.