Hanna Instruments HI 8633 User Manual

Page 10

background image

10

• Press and hold down

Temp to display the

temperature.

• Adjust the TEMPERATURE knob to display 20°C.

• Release

Temp key to display conductivity measurement.

• Select 19.99 mS/cm (

HI 8633)

or 1999 mg/L (

HI 8734) range

by pressing the appropriate
range key.

• Follow LCD calibration indication. Adjust the calibration knob until

the display shows for

HI 8633 the conductivity reading at 25°C

(see the conductivity vs. temperature table), e.g. @ 25°C,
12880

µ

S/cm = 12.88 mS/cm, or for

HI 8734 the TDS

reading at 25°C (see the TDS vs. temperature chart), e.g.
@25°C, 1382 mg/L.

• All subsequent measurements will be compensated to 25°C (77°F).

If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE
knob at 18°C (if the temperature of the solution is 18°C), adjust
the knob to read “

11.67 mS” (see the conductivity vs. tempera-

ture chart) or “

1251 mg/L” (see the TDS vs. temperature chart). All

subsequent measurements will be compensated to 20°C.

• The calibration is now complete and the instrument is ready for

use.