pH ELECTRODE CLEANING PROCEDURE
• General

Soak in Hanna HI 7061 or HI 8061 General

Cleaning Solution for approximately ½ hour.

• Protein

Soak in Hanna HI 7073 or HI 8073 Protein

Cleaning Solution for 15 minutes.

• Inorganic

Soak in Hanna HI 7074 Inorganic Cleaning

Solution for 15 minutes.

• Oil/grease

Rinse with Hanna HI 7077 or HI 8077 Oil and

Fat Cleaning Solution.

IMPORTANT: After performing any of the cleaning procedures, rinse the

electrode thoroughly with distilled water, refill the reference chamber with

fresh electrolyte (not necessary for gel-filled electrodes) and soak the electrode

in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking

measurements.

EC Probe Maintenance
Rinse the probe with clean water after measurements. If a more thorough

cleaning is required, remove the probe sleeve and clean the probe with a cloth

or a nonabrasive detergent. Make sure to reinsert the sleeve onto the probe

properly and in the right direction. After cleaning the probe, recalibrate the

instrument.

The platinum rings support is made of glass. Take great care while handling

the probe.