EC/TDS CALIBRATION

EC calibration is a one-point procedure. Selectable calibration points are 0.00 µS

for offset and 84.0 µS, 1413 µS, 5.00 mS, 12.88 mS, 80.0 mS, 111.8 mS for

slope.

Rinse the probe with calibration solution or deionized water.

Immerse the probe into the solution. The sleeve holes must be

completely submerged. Tap the probe repeatedly to remove

any air bubbles that may be trapped inside the sleeve.

To enter EC calibration, select the EC range and press CAL.

Note: TDS reading is automatically derived from the EC

reading and no specific calibration for TDS is needed.

Pressing CAL when TDS range is selected has no

effect.

For zero calibration, just leave the dry probe in the air. This calibration is

performed in order to correct the reading around 0.00 µS point. The slope is

evaluated when the calibration is performed in any other point.

The “BUF” and “CAL” tags are displayed. The primary LCD will display the EC

reading. The secondary LCD will display the buffer value. The “ ” and “~”

tags will blink.

Select the desired value with the ARROW keys, if necessary.

When the reading is stable, “READY” tag is displayed and “CFM” tag starts

blinking on the LCD, asking for confirmation.

Press CFM to confirm calibration.

The instrument stores the calibration value and returns to

measurement mode.
Note:

If the uncalibrated reading is too far from the expected value, the

“WRONG” and “ ” tags will blink. Calibration can not be confirmed.

In this case check if the correct buffer has been used or clean the probe

by following the Cleaning Procedure (see page 45).

S

BUF

CAL

S

BUF

CAL

READY

CFM